background image

i

MN1-6031 Rev. 3

C22T Probe

Instruction Manual

Specification

MN1-6031 Rev. 3

Notes for operators and responsible maintenance personnel

★ 

Please read through this Instruction Manual as well as the separate Instruction 
Manual “Safety (MN1-5984)” and “Cleaning, Disinfection and Sterilization 
(MN1-6000)“ carefully prior to use.

★ 

Keep this Instruction Manual together with the ultrasound diagnostic instru-
ment for any future reference. 

© Hitachi, Ltd.  2015, 2017. All rights reserved.

Summary of Contents for C22T

Page 1: ...nnel Please read through this Instruction Manual as well as the separate Instruction Manual Safety MN1 5984 and Cleaning Disinfection and Sterilization MN1 6000 carefully prior to use Keep this Instruction Manual together with the ultrasound diagnostic instru ment for any future reference Hitachi Ltd 2015 2017 All rights reserved ...

Page 2: ...ii MN1 6031 Rev 3 ...

Page 3: ...prevent hazards and injuries to the operator or patients Danger Indicates an imminently hazardous situation which if not avoided will result in death or serious injury to the operator or patient Warning Indicates a potentially hazardous situation which if not avoided could result in death or serious injury to the operator or patient Caution Indicates a potentially hazardous situation which if not ...

Page 4: ...rmation 1 1 1 Intended use 1 1 2 Classification of ME equipment 1 1 3 Standard components 1 1 4 Option 1 2 Specifications and Parts name 3 2 1 Specifications 3 2 2 Name of each parts 4 3 Preparations before use 5 3 1 Visual check 5 3 2 Confirmation of cleaning and sterilization 5 3 3 Operation check 5 4 Operation 7 4 1 Operation 7 ...

Page 5: ...range of IPX7 Classification based on the degree of protection against electric shock Type BF applied part Classification for protection against ingress of liquids IPX7 Watertight equipment Operation mode Continuous operation Method of sterilization Refer to the separate instruction manual Cleaning Disinfection and Sterilization 1 3 Standard components The standard components of C22T probe are as ...

Page 6: ... 2 MN1 6031 Rev 3 ...

Page 7: ... of view 82 Frequency 3 5 MHz Cable length 2 9 m Service life 3 years Applied part Probe tip including ultrasonic radiation part see the section 2 2 Part treated as applied part Cable up to 0 2 m length from the probe tip IPX7 range See Figure 1 In case that not putting the waterproof case to the ultrasound probe connector In case that putting the waterproof case to the ultrasound probe connector ...

Page 8: ... corresponds to the side of the orientation mark on the image Probe tip This part is gripped during operation Cable Cable transfers electric input output signals Connector The connector is the part which is connected to the ultrasound diagnostic instrument Caution Do not pull bend twist or apply excessive force to the cable The probe may malfunction due to cable disconnection Do not subject the ul...

Page 9: ...ct the probe to the ultrasound diagnostic instrument and check that the displayed scan type and frequency correspond to those of the probe Check also that there is no abnormality in the image Remark Please refer to the documentation supplied with the ultrasound diagnostic instrument for how to connect the probe and information displayed on the monitor If the probe is operated in still air brightne...

Page 10: ... 6 MN1 6031 Rev 3 ...

Page 11: ...r to the documentation supplied with the ultrasound diagnostic instrument Caution Do not operate the probe with excessive force Use with excessive force could result in injury to the patient Scan for minimum time necessary at the lowest possible acoustic output Acoustic output may affect the patient s internal tissues For details about the acoustic output please refer to the documentation supplied...

Page 12: ...08 31 Manufacturer Hitachi Ltd 2 16 1 Higashi Ueno Taito ku Tokyo 110 0015 Japan Contact 81 3 6284 3668 http www hitachi com businesses healthcare index html Overseas Offices Hitachi Medical Systems GmbH Otto von Guericke Ring 3 D 65205 Wiesbaden Germany EU Importer Hitachi Medical Systems Europe Holding AG Address Sumpfstrasse 13 CH 6300 Zug Switzerland ...

Reviews: