QUICK START

TM

AP 720S   Semi-Automated Instrument

Note:

 The assumed starting point is the instrument state after 

completion of a previous run and following routine maintenance and 

shutdown. 

(The alternate state would be the interim between two

successive runs of the instrument)

. The wash buffer and deionized 

water tubes will be on a clean surface following completion of the 

drain function and the power will be off.

Prepare Serum Samples

1.

Centrifuge all serum samples.

2.

Transfer each serum sample into an appropriately labeled clean 

plastic storage tube with lid.

Load Sample / Reagent Rack

3.

Load the required number of empty Sample Cups into the 

Sample / Reagent Rack. Start at position #1 and do not leave 

any empty positions.

4.

Transfer the correct volume of each prepared sample into the 

sample cups:

600

µ

l serum for a 36-allergen panel

490

µ

l serum for a 20-allergen panel

Note:

 The red Sample Fill Line on the Sample / Reagent Rack can be 

used as a guide to check that there is a fill volume of approximately 

600

µ

l in the Sample Cups.

5.

Load the Antibody Reservoir (BLUE line) onto the Sample / 

Reagent Rack.

6.

Remove the Antibody Reagent bottle from the kit and gently 

invert prior to use.

7.

Fill the Antibody Reservoir with the required amount of reagent

as listed in the Instruction Manual.

8.

Install the Sample / Reagent Rack on the instrument aligning 

sample position #1 with the red indicator line and lowering the 

rack so that the Center and Positioning Pin locate into the holes 

at the bottom of the rack.

Load Pette Rack

9.

Remove the required number of OPTIGEN Pettes from the kit 

box and label with sample identification. (For a given run all 

pettes must be of the same panel type, see Step 21 below). 

Reseal the plastic bag and return the kit to the refrigerator.

10.

Attach a Pette Tip to each Pette Body.

11.

Place the Pette Rack on the Pette Stand. (Always use the Pette 

Stand when loading or removing Pettes.)

12.

Load the Pettes into the Pette Rack. Always start at position 

#1 and do not leave any empty positions. Check that each Pette 

Body is correctly secured in the Pette Rack. Ensure the number 

of Pettes matches the number of samples.

13.

Remove the Pette Rack from the Pette Stand.

14.

Install the Pette Rack on the instrument by lowering it onto 

the center guide bar with pette cartridge #1 positioned front 

and center so that the rack is properly aligned with the 

positioning pin.

15.

Close the door of the instrument.

Load Wash Buffer and Deionized Water, 

and Empty Waste

16.

Prepare the Wash Buffer according to the instructions in the 

OPTIGEN package insert.

17.

Fill the Wash Buffer Bottle with the required amount of Wash

Buffer as listed in the Instruction Manual.

Secure the yellow cap on the Wash Buffer Bottle and ensure the 

weighted end of the Wash Buffer tube is located at the bottom 

of the bottle.

18.

Ensure there is sufficient (1L) deionized water in the Deionized 

Water Bottle and fill if needed. Secure the white cap on the 

Deionized Water Bottle and ensure the weighted end of the 

Deionized Water tube is located at the bottom of the bottle.

19.

Empty the Waste Bottle by pressing the quick release connectors 

to detach the tubes, unscrew the cap, and then empty the waste 

liquid. Reconnect the Waste Bottle by screwing the cap back on 

and tightening. Attach the quick release connectors to the cap, 

matching the color-coded labels.

Doc. No. 0822 - ENG Rev. 00

www.hcdiagnostics.com